The LC-MS/MS analysis of environmental contaminants in water and sediments upon negative ESI
The LC-MS/MS analysis of environmental contaminants in water and sediments upon negative ESI
Here is demonstrated the analysis of some acidic environmental contaminants that can be determined in water or sediment samples in our laboratory. For the analysis of water samples, there was used a combination of large-volume injections with highly sensitive MRM acquisition. The large-volume injections save time on sample preparation and provide a solution for recovery-related problems sometimes observed using SPE preconcentration techniques. For improving the accuracy of results, the samples are spiked with available isotopically labeled internal standards.
For the analysis of organic contaminants in sediments, the samples are extracted with organic solvents using an accelerated solvent extractor (ASE, Dionex).
Instrument: QTRAP 6500+ mass spectrometer (Sciex) coupled with Nexera binary UPLC (Shimadzu). The mass spectrometer was operated in negative ESI scheduled MRM mode. Chromatographic separations were carried out using the Luna Omega Polar-C18 column (2.1×150 mm, 3 µm particle size, Phenomenex) using a linear gradient of ACN/water (with 0.1% acetic acid). The instrument was controlled and data were analyzed using Analyst and MultiQuant software respectively (Sciex).
Negative ESI LC-MS MRM chromatograms of environmental contaminants (concentration of analytes, - 50 pg/ml).
A list of MRM parameters for environmental contaminants determined in negative ESI mode.